UV Power Puck? FLASH -Operating the Radiometer
Turning “ON” the Radiometer: Press and Hold the ON / OFF button until the display illuminates. When the button is
released, the display will briefly display the Radiometer Model Name, Serial Number, Software Version, Calibration
Date, Range, and Wavelength Bands installed. The display will then enter the default mode and display the data from
the last run before the unit was turned off. Continuing to hold down the button at startup will extend the display time
until released.
Turning “OFF” the Radiometer: Press and Hold the ON / OFF button. A tone will sound. When tone stops, release
the button. The unit turns off.
Entering the “RUN” MODE: A short press of the “RUN” button clears the memory and puts the unit in the “RUN”
mode. The display shows “RUNNING” after shortly displaying the internal temperature of the unit. Confirm that the
unit displays “RUNNING” before initiating a reading.
Place the radiometer on the belt or object with the optic window looking toward the UV source. The display and
buttons will be facing away from the UV source. When the radiometer exits the curing chamber, the display will still
be flashing “RUNNING”.
CAUTION: Exposing the display to high UV radiation will damage the display.
Exiting the “RUN” MODE: A short press of the “STOP” button (Soft button display bar indicates “STOP” next to the
“ON / OFF” button) will exit the “RUN” mode and will return to the same default mode prior to making the exposure
run, but will display the new value.
MODE:
ALL CHANNEL – displays Joules reading for each of the four (4) UV bands (UVA, UVB, UVC, UVV)
REFERENCE – Make a run, data will appear next to the UV band. CAUTION: Be sure you want to
overwrite the current data on the REF line before pressing SAVE. Press SAVE, data is transferred to the
REF line. The data will remain until it is overwritten. The difference or change between the current run data
and the reference data is displayed as a percentage change on the %DIFF line.
GRAPH – Illustrates the irradiance profile for the UV source(s). UV data is stored and displayed as a graph of
time (X axis) vs. intensity (Y axis) for each UV lamp source.
SMOOTH: n/a (Not applicable)
UNITS: J, mJ, μJ – Select the unit values
DISPLAYS: LOW, MEDIUM, HIGH – Select the display intensity
SAVE D ? EXIT – Soft Button Indicators