    4. Install the free end of the adapter tubing onto the calibration adapter.

    5. Press the calibration adapter onto the BioStage inlet until a fi rm seal is established.

    6. Connect the BioStage outlet to one end of the ?-inch ID tubing.

    7. Insert the small end of a second Luer adapter into the free end of the ?-inch

    tubing.

    8. Insert the large end of the second Luer adapter into one end of the 3/8-inch

    sampling tubing and the pump inlet adapter into the other end.

    9. Insert the pump inlet adapter into the pump inlet.

    10. Calibrate the sample pump to 28.3 L/min for a Standard BioStage or

    14.15 L/min for BioStage 200 (see calibration procedure in QuickTake 30 operating

    instructions 40079).

    11. When calibration is complete, remove the calibration adapter and tubing and

    fl owmeter.

    The calibration adapter and rotameter are for use with both models of BioStage. See Accessories/

    Replacement Parts.

    Sample

    Read the following before sampling.

    ? Sanitize hands or wear disposable gloves. Sanitize hands and clean the BioStage

    Impactor in between samples.

    ? Never use agar that has expired, displays visible cracks, or has been contaminated.

    ? Allow agar plates to warm up to ambient temperature (approximately 20 minutes)

    before use.

    ? When sampling indoors, close all doors and windows that could affect airfl ow in the

    sampling area.

    1. The BioStage outlet should be connected to

    the pump inlet with fl exible tubing or using

    the Mounting Bracket accessory (Cat. No.

    228-9531); see Operating Instructions 40066

    supplied with that accessory (see right).

    2. Remove the inlet cone and jet classifi cation

    stage from the BioStage. Place a new,

    unexposed agar plate into the base plate (see

    Assemble with Media). Remove the lid from

    the agar plate and place in a clean resealable

    bag. Reassemble the impactor.

    3. Turn on the pump and sample for 2 to 5 minutes.

    Sampling for too long can cause overgrowth of sample. Sampling times that are

    too short can cause false negatives.

    我們營業(yè)的時(shí)間9:00-18:00

    • 技術(shù)
    關(guān)閉在線客服

    Nano電子商城微信公眾號(hào)手機(jī)商城 Nano電子商城微信公眾號(hào)二維碼

    手機(jī)商城           微信商城

    任何購物問題請(qǐng)聯(lián)系納米科技(香港)有限公司  | 江蘇省蘇州市昆山開發(fā)區(qū)章基路189號(hào)科創(chuàng)基地2號(hào)樓202室 Tel:021-52069907 email: 19mro@19mro.com

    工作時(shí)間:周一至周五 8:00-18:00; 網(wǎng)站備案號(hào) :滬ICP備12049383號(hào); Copyright?2009-2012 19mro.com 版權(quán)所有 


     
    亚洲A∨无码无在线观看| 日日摸日日碰夜夜爽无码| 未满小14洗澡无码视频网站| 久本草在线中文字幕亚洲欧美| 久久亚洲AV成人无码| 无码精品A∨在线观看中文| 亚洲av无码无在线观看红杏| 日本中文一区二区三区亚洲| 亚洲AV永久无码精品一区二区 | 欧美中文字幕在线| 蜜臀av无码人妻精品| 人妻丰满熟妇岳AV无码区HD| 精品高潮呻吟99av无码视频| 最近新中文字幕大全高清| 超清中文乱码字幕在线观看| 国产av永久无码天堂影院| 99精品人妻无码专区在线视频区| 中文字幕亚洲色图| 痴汉中文字幕视频一区| 人妻无码人妻有码中文字幕| 亚洲AV永久无码精品一区二区| 97免费人妻无码视频| 狠狠躁天天躁无码中文字幕 | 99无码人妻一区二区三区免费| 亚洲熟妇无码八AV在线播放| 亚洲七七久久精品中文国产| xx中文字幕乱偷avxx| 最近2019中文字幕免费大全5| 欧美日韩中文字幕2020| 性色欲网站人妻丰满中文久久不卡| 亚洲中文字幕日本无线码| 色噜噜亚洲精品中文字幕 | 中文字幕在线免费观看| 欧美日韩国产中文字幕| 国产中文字幕在线免费观看| 最近中文字幕完整在线看一| 一二三四在线观看免费中文在线观看 | 久久亚洲精品无码aⅴ大香| 中文字幕无码第1页| 亚洲av无码无在线观看红杏| 久久久久久亚洲AV无码专区|