Standard maintenance and calibration service for the Condensation Particle Counter Model 3776 consists
of these procedures:
1) Review written instructions or explanation of problem from customer
2) Initial inspection
3) Update manual, software and firmware as necessary
4) Clean instrument and optics
5) Inspect and replace as needed: wick, filters, tubing, and O-rings
6) Leak check
7) Verify flow and temperature control
8) Align optics
9) Zero check
10) Final electrical checkout and performance verification
Standard service includes minor repairs and replacement parts. Major repairs and replacement parts will
be charged for parts and labor. TSI will contact the customer for approval before making any major
repairs.
Turnaround time for standard service is approximately two weeks after receipt of the instrument. TSI
service work is backed by a 90 day service warranty.